Chapters Transcript Setting Up and Operating the Kangaroo™ ePump for Feeding and Flushing Learn how to set up and operate the Kangaroo™ ePump when using a feeding set with flush bag. The first step of basic pump operation is to prepare and load your feeding set. To prepare your feeding set, simply fill the feeding bag or spike the feeding container per your standard procedure. In order to achieve proper accuracy. The bottom of the feeding setback must be 18 inches above the top of the feeding pump. Caution failure toe Hang the set to the proper level can result in inaccurate system feeding. Once you're feeding. Set is ready. Press the power button on the lower right of the front panel to power up pump. You have the option to either keep your previous settings or clear settings to keep settings. Option will allow you to keep your previously program settings clear. Settings option will clear your previous settings and the volume fed from your previous feeding. Next opened the blue door to the feeding set loading area. Grasp the thumb tab on your feeding set and insert the tap directly into the left pocket of the loading set area. Press down firmly on the thumb tab and aligned with white line to ensure it is seated. Graphs the black green retainer and gently wrapped the tubing counter clockwise around the road. Er, carefully pull the black retainer end upward to position it over the slot on the right and then lower the retainer into the slot. Use caution not to overstretch the tubing. Close the blue door to the feeding set loading area it pumps. Setback that has been emptied will trigger the feed era screen. In this condition, the pump setback can be refilled to continue the feeding, but only after the pump set has been re primed. To re prime the pump. Set first disconnect the feeding line to the patient and then refill the bag. Then press. Continue to begin running the pump household press, adjust settings, press prime pump press. Hold the prime to prime the line and press done to return to the main. Many of the resume in feature is similar to the keep tube open feature and that it allows for inactive time. Select the resume in feature and the feeding pump will pause the delivery of feeding nutrition for pre programmed time of 30 minutes. To activate the resume In feature, simply press hold during the feeding cycle and select resume in the pump display will show resume in 30 minutes used the adjust settings feature to change any programming component during a feeding. To use this feature, select hold while the pump is running. Select Adjust settings to adjust all settings. Select run to return to normal operations. Programming your pump for continuous modus Simple. The ATM style interface uses on screen prompts and arrow screens to set the flush volume and flush interval, as well as the speed of feed delivery and, if desired, the total volume to be delivered. To begin, select the adjust flush option in the opening menu. Press flush volume to define the volume per flush occurrence. Press the buttons to program the flush volume from 10 to 500 increments of one million later. Separate arrows control the one hundreds tens and once once you reach the desired flush volume, press enter to exit press flush interval to define the time interval between the beginnings of each flush. Press the arrow buttons to program the time interval from 1 to 24 hours and increments of one hour. Once you reach the desired interval, press enter to Exit and Preston to return to the main menu. Select the adjust feed option in the main menu press feed rate. To define the rate of delivery, press the buttons to program the rate from 1 to 400 increments of one millimeter per hour. Separate arrows control the one hundreds, tens and ones. Once you reach the desired feed rate, press enter to exit the menu. If you'd like to set the total volume to be delivered, select fee TBD and input the total volume to be delivered. Press the buttons to select the feet. Volume Note. Volume to be delivered is an optional feature Onley. Use the V TBD if you want the pump to stop and alarm once the set amount of formula has been delivered, caution. When you're using the V TBD feature, do not turn the pump off during the feeding. This will result in an internal resetting with the volume to be delivered. Please use the hold resume in or Katie Oh features as referenced and advanced features. Press enter to exit programming your pump for intermittent modus simple. The ATM style interface uses on screen prompts and arrow screens to program the bullets rate, the number of Pulis's, the volume per Bullis and the bullets interval to begin select theaters. Just flush option in the opening menu. Press flush volume To define the volume per flush occurrence. Use the arrow buttons to program the flush volume from 10 to 500 increments of one millimeter. Separate arrows control the one hundreds tens and ones. Once you reach the desired flush volume, press enter to exit. Press flush interval to define the time interval between the beginnings of each flush. Press the arrow buttons to program the time interval from 1 to 24 hours and increments of one hour. Once you reach the desired interval, press enter to Exit and Preston to return to the main menu. Next select Adjust Bullis in the opening menu. Press Bullis rate. To define the rate of delivery, press the arrow buttons to program the delivery rate from 1 to 400 increments of one millimeter per hour. Separate arrows control the one hundreds tens and once once you reach the desired rate of delivery, press enter to exit press number of Pulis's To define the total number of Pulis's to be delivered. Use the Arrow buttons to program the number of Pulis's up to a maximum of 99 the number of Pulis's can be set at zero, in which case the pump will run until the bag is empty. Once you reach the desired number, press enter to exit press volume per Bullis to define the volume to be delivered in each Bullis. Press the buttons to program the volume. Purple is from 1 to 1000 millimeters in increments of 1 mL. Once you reach the desired volume, press enter to exit. Finally press Bullis interval to define the time interval between the beginnings of bullets deliveries. Press the buttons to program the time interval between starts of Bullis deliveries from 1 to 24 hours and increments of one hour. Once you reach the desired interval, press enter to exit Preston to return to the home screen and then select run to begin operating the pump he keep to open or Katie Oh feature. When activated, keeps the rotor of the pump moving at extremely low rates. The small amount of nutrition delivered ensures the Peyton C of the tube is maintained. Katie Oh option will appear on the screen a few seconds after the run button is pressed. To use this feature, simply press the K T o button. Note. The K T O feature could be programmed from 5 to 240 minutes. Try setting it to 60 minutes at the end of the program. Time the pump will resume. It's normal operation. During the running mode, the pump can be diverted to immediately perform a flush on demand. Press flush now in the running menu. Press the buttons to program the volume to be immediately flushed. Press enter to exit the menu, setting the flush. Now volume is temporary and will not change the main flushing volume previously. Program. The flush now is accidentally pressed. The screen will return to the running screen after 10 seconds, provided that no input for the flush volume is entered. Note. Flush now is Onley available in feed and flush mode. Created by