Chapters Transcript Pump Status Overview In the upper right corner of the kangaroo Omni Enteral feeding pump. The pump status indicators are led lights that provide a quick visual indication of the pump status. There are five different light patterns. First, a solid green light means the pump is running normally. Second, a blinking green light indicates the pump is ready to provide nutrition. Third, a solid yellow light is a low priority alarm. The pump is still running normally but it will need the user to take action soon or else it will stop functioning properly. Fourth, a blinking yellow light indicates a medium priority alarm. The pump has stopped delivering nutrition and the user needs to take action for the pump to run normally again. Fifth, a blinking red light indicates a high priority alarm warning. The pump has stopped delivering nutrition and will be out of service for a period of time. For more information, please refer to the kangaroo Omni Enteral feeding pump instructions for use. Created by